How to convert a Fluke 77 IV average responding multimeter into a True RMS model 177 for 10 cents.
Should also work for converting a Fluke 83V to 87V
True RMS measurement vs Average Responding
And calibration of a Fluke meter.

00:00 – Hacking a Fluke 77 IV Multimeter
00:35 – Is the Fluke 70 series discontinued?
01:15 – Fluke History
03:10 – I noticed something…
04:49 – The AD737 True RMS converter chip
06:04 – Theory of Operation
07:21 – What is Crest factor
08:58 – How do they do it?
09:25 – Fluke 87V Schematic and the custom ASIC
10:53 – Let’s modify it!
12:58 – A sine wave is NOT a crest factor of 1
14:01 – Soldering
15:15 – Will it work?
16:46 – Calibration procedure
20:09 – Winner Winner Chicken Dinner